用于分析低浓度气体的具有长路径长度池的真空FTIR

收藏
检测样品: 工业气体
检测项目: 低浓度气体分析、CO、CO2、NO、NO2
浏览次数: 55
发布时间: 2023-07-06
关联设备: 1种 查看全部
获取电话
留言咨询
方案下载

佳士科商贸有限公司

银牌9年

解决方案总数: 276 方案总浏览次数:
方案详情
关键词:真空FTIR长路径气室,气体分析,一氧化碳二氧化氮,一氧化氮二氧化碳CO,NO,NO2CO2

方案详情

低浓度气体或蒸气化合物的FTIR分析通常需要高S/N测量。然而,由于仪器的光路中存在大气水蒸气和二氧化碳,即使使用氮气吹扫,一些气体的测量也可能很困难。分析NO特别困难,因为吸收带与水蒸气在同一光谱区域,测量NO2、CO2和CO气体也很困难,它们的吸收带与CO2吸收在同一区域。使用全真空光谱仪,可以完全消除光路中的水蒸气和二氧化碳,并且可以在低得多的浓度下进行测量。Figure 1 Spectrum of a 2 ppm CO sample in a 20 m gas cell, with a S/N approx. 200:1. Figure 2 The noise level of a  spectrum measured with the same conditions is ~4 x 10-5 Abs. At this noise level, measurement of CO gas can be easily  made with concentrations as low as 20-50 ppb.‍Application Note230TR0165-E Vacuum FTIR with Long Path Length Cell for theAnalysis of Low Concentration Gases2/2Application Note Vacuum FTIR with Long Path Length Cell for the Analysis of Low Concentration Gases Introdu c tio n A high S/N measureme n t is generally required for FT I R ana l ysis of low concentrat i ons of gas or vapor com p o u nds.Howeve r the measurement of so m e gases can be di f ficult due to the presence of atmospheric water vapor and ca r bon d i ox i de present in the optical path of the instrument , even wit h n itrogen purging. It is especial l y d if ficul t to analyze NO because the absorption ba n d i s in the same spectra l region a s water vapor, it is also d if ficul t to measure NO,,CO,and CO gases, whose absorption bands are i n the same region as CO, absorpt i on. Usi n g a full-vacuum spectrometer, water vapor a n d carbon dioxide i n the light path can be completely el im inated and the measurement can be made at mu c h l ower concentr a tions. Figure 1 Spect r um of a 2 ppm CO sample in a 20 m gas cel l , wit h a S/N approx. 200:1. Figu r e 2 The n oise level of a spectrum measured with the same condit i ons is ~4x 10-5 Abs. At t h is noise level , measurement of CO gas can be easily made with concentrations as low as 20-50 ppb. Keywo r ds Vacu u m FTIR long path gas cell, gas analysis, ca r bon monoxide Nitrogen d i ox i de, nitr i c ox i de carbo n dioxide CO, NO,NO,CO. Condition R e so lu ti o n : 4c m -1 Sca n s : 500 Detector : MC T C ell: 2800 2600 Wavenumber[cm-11
确定

还剩1页未读,是否继续阅读?

不看了,直接下载
继续免费阅读全文

该文件无法预览

请直接下载查看

产品配置单

佳士科商贸有限公司为您提供《用于分析低浓度气体的具有长路径长度池的真空FTIR》,该方案主要用于工业气体中低浓度气体分析、CO、CO2、NO、NO2检测,参考标准--,《用于分析低浓度气体的具有长路径长度池的真空FTIR》用到的仪器有JASCO傅立叶变换红外光谱仪FT/IR-6000