Kyoto Electronics Manufacturing Co.,Ltd.Data No.LN38-090427 No.EKCR-090427 KEM Application Note No.EKCR-090427 Coulometric Karl-Fischer Titration Moisture of N-methyl-2-pyrrolidone ·Summary- Checked the possibility of measuring moisture of N-methyl-2-pyrrolidone with MKC-520 andgood repeatability was obtained. -Measuring instrument- Main unit: IMKC-520 Coulometric KF moisture titrator Electrode:M-713 Twin platinum electrode #433-0006Inner burette -Reagent- Anolyte: Coulomat AK Catholyte:Coulomat CG-K -Measurement method- Preparation: (1) Fill the titration cell with the anolyte and the inner burette with the catholyte.(2) Pretitrate for dehydration. Measurement: (1) Place 0.1g sample in the titration cell with syringe for the moisture measurement on thesample. -Equation- Moisture (%): ((Data-Driftxt-Blank)/(Wt1-Wt2))xF Data: Measured water content (ug) Drift : Drift before measurement (ug/s) t: Measuring time(s) Blank : Blank value (ug) Wt1: Tare including sample weight (g) Wt2: Tare including residual weight (g) F :Factor(1.00) -Standard method- Karl Fischer method .Ambient conditions- [Titration1 Method 2 t(stir) 0s t(wait) 15s t(max) 0s Drift Stop : Rel 0.1 ug/s Cont. Gain 5.0 Stable 0.1 ug/min Start Manual Oven : Off 「Calculatio n] Clac. No. 2 Unit : Weight :\Variable Drift Comp. : Auto 「DataList1 Data Samp. Time 30s Display : Off Print Meas. Data: Off Print Graphic : Off [Result] No, wgH20 Conc[a 01 823.2 0.6697 02 840.8 0:6605 03 767.0 .6618
Date 2009/04/27 14:59Semple No.