卡尔费休库仑法微量水分测定仪测定丁酮中水分含量

收藏
检测样品: 其他
检测项目: --
浏览次数: 1928
发布时间: 2009-04-29
关联设备: 4种 查看全部
获取电话
留言咨询
方案下载
方案详情
卡尔费休库仑法微量水分测定仪测定丁酮(MEK)中水分含量 使用日本京都电子公司(KEM)-卡尔费休库仑法微量水分测定仪(MKC-520),测定丁酮(MEK)中水分含量的应用资料。

方案详情

Kyoto Electronics Manufacturing Co., Ltd.Data No.LN38-090429上海市漕溪路222号航天大厦北楼902室网址: http://www.kyoto-kem.net.cn电话:021-54488867 电邮: kemu-kem@163. com KEM Application Note No.EKCR-090429Coulometric Karl-Fischer Titration Moisture of Methyl ethyl ketone ·Summary- Checked the possibility of measuring moisture of Methyl ethyl ketone with MKC-520 andgood repeatability was obtained. -Measuring instrument- Main unit: MKC-520 Coulometric KF moisture titrator Electrode: M-713 Twin platinum electrode #433-0006Inner burette -Reagent- Anolyte: Coulomat AK Catholyte: Coulomat CG-K -Measurement method- Preparation: (1) Fill the titration cell with the anolyte and the inner burette with the catholyte. (2) Pretitrate for dehydration. Measurement: (1) Place about 0.1g sample in the titration cell with syringe for the moisture measurement onthe sample. -Equation- Moisture (%): ((Data-Driftxt-Blank)/(Wt1-Wt2))xF Data : Measured water content (ug) Drift : Drift before measurement (ug/s)t:Measuring time(s) Blank : Blank value (ug) Wt1 : Tare including sample weight (g) Wt2 : Tare including residual weight (g) F : Factor (1.00) ·Standard method- Karl Fischer method -Ambient conditions- Temperature: 26°℃ R. Humidity: 42 % Weather: Fine 京都电子工业株式会社-可睦电子(上海)商贸有限公司 Titration] Method 2 t(stir) 0s t(wait) 15s t(max) 0s Drift Stop : IRel 0.1 ug/s Cont. Gain 5.0 Stable 0.1 ug/min Start :1Manual Oven :Off 「Calculation] Clac. No. 2 Unit Weight :%:\Variable Drift Comp.: Auto [Data Listl Data Samp. Time 30 sDisplay : Off Print Meas.Data: Off Print Graphic : Off Resy.lttX* Date 09/04/2912:38 SamPle No. 01-02WI 0.07339W12 0 Net 0.0733 Result 107.1L19 0.1461% 关关Resi Ii E关关关 Date 09/04/29 12:39 Samele No. 01-03It1 0.0969寸+2 0gNet 0.09699Resu1t 138.6 u0.1430 **XResul t 光光 Date 09/D4/2912:40 Samele No, 01-04Wtl 0.12379W+2 09Net 0.1237 Result 178.8US D.1445 -Measurement results- Run Sample size(g) Moisture(ug) Concentration(%) Statistics 123 0.0733 0.0969 0.1237 107.1138.6 178.8 0.14610.1430 0.1445 MeanSD RSD 0.1445%0.0016% 1.0726% -Comments- -Comments- 京都电子工业株式会社-可睦电子(上海)商贸有限公司电话:021-54488867 电邮: kemu-kem@163.com 上海市漕溪路222号航天大厦北楼902室网址: http://www.kyoto-kem.net.cn
确定

还剩1页未读,是否继续阅读?

不看了,直接下载
继续免费阅读全文

该文件无法预览

请直接下载查看

可睦电子(上海)商贸有限公司-日本京都电子(KEM)为您提供《卡尔费休库仑法微量水分测定仪测定丁酮中水分含量》,该方案主要用于其他中--检测,参考标准--,《卡尔费休库仑法微量水分测定仪测定丁酮中水分含量》用到的仪器有MKH-710M容量法库仑法混合型卡尔费休水分仪、MKC-710B卡尔费休库仑法微量水分测定仪、MKC-710S豪华型库仑法卡尔费休水分仪、MKC-710M旗舰型库仑法卡尔费休水分仪