GC 柱温箱加速插入套件-用于Agilent 6890和7890仪器



  • Get the same GC separation in less time—use a GC Accelerator kit and the EZGC method translator to accurately convert methods to a scaled-down column format.

  • Scaled-down methods let you speed up analysis time and increase sample throughput without capital investment.

  • GC Accelerator kit installs easily without damaging the GC column or interfering with the MS interface.

Designed with GC-MS users in mind, the GC Accelerator kit provides a simple way to speed up sample analysis. By reducing oven volume, these inserts allow faster ramp rates to be attained, which reduces oven cycle time and allows for increased sample throughput and more capacity to process rush samples. When faster ramp rates are used, existing methods can be accurately scaled down to smaller, high-efficiency, narrow-bore columns using Restek’s EZGC method translator. With a scaled-down column, a properly translated method, and a GC Accelerator kit, you can obtain the same chromatographic separation—often with greater sensitivity—in a fraction of the time without making a capital investment.


订货信息:

GC 柱温箱加速插入套件-用于Agilent 6890和7890仪器

描述货号
GC 柱温箱加速插入套件23849


GC Accelerator Use

Installation

  1. Turn off the oven and wait until the oven, inlets, detectors, and MS transfer line have cooled.

  2. Ensure the column and the column hanger are in the back position in the oven

  3. If no detector is in the front position, install the two block inserts in front of the column as shown in Figure 2. If there is a detector in the front position, install the two block inserts in front of the column as shown in Figure 3.

Caution: The block inserts should not touch the analytical column or the MSD source nut (Figure 4).

4. Install the plate insert in the front of the oven so that the front is flush with the edge of the oven (Figure 5).


Removal

Turn off the oven and wait until all oven, inlets, detectors, and MS transfer line surfaces have cooled.

Caution: To avoid burns, use the strap to remove the plate insert or wear heat-resistant gloves. The metal fasteners and label tag on the plate insert may still be hot even after the GC oven has cooled down.

  1. Grip the strap on the plate insert and remove it by pulling the bottom edge out first (Figure 6).

  2. 2. Remove the block inserts carefully; do not allow them to contact the column or the MSD source nut.